CERTIFICATE OF ANALYSIS
Catalog No.: |
30907YB |
Quantity: |
Not less than 1mL |
Product name: |
Aromatics-Mix2 |
Lot No. |
20245-016-1 |
Concentration: |
Each100ug/mL(±3%,k=2) |
Expiry Date: |
2023-02-18 |
Solvent: |
Cyclohexane/Acetone(1:1) |
Storage: |
0-8℃ and dark |
|
Component |
CAS |
Conc.(ug/mL) |
Purity |
Formula |
M (g/Mol) |
1 |
2456-Tetrachloro-m-xylene |
877-09-8 |
1000 |
97.0% |
C8H6CI4 |
243.94 |
2 |
PCB 209 |
2051-24-3 |
1000 |
98.0% |
C12CI10 |
498.66 |
Instructions for handing and correct use:
The RMshould be equilibrated to room temperature and shaken prior to use.The concentration with its uncertainty is guaranteed in the range between(20±4)℃.The expiry date is valid for original sealed bottles under recommended storage conditions only.Open slowly and carefully to avoid dispersion ofthe material
Intended use:
For laboratory analysis only not for drug, household or other uses.
Uncertainty:
The reported uncertainty is the expanded uncertainty estimated in accordance with ISO 17034 with a coverage factor k=2,corresponding to a level of confidence ofabout 95%.
Measurement traceability:
Traceability to SI has been established through an unbroken chain ofcomparisons each having stated uncertainties. All analytical balances and Class A glassware used in manufacturing and quality control are calibrated by an accredited calibration laboratory.
Safetyinformation:
Treat as hazardous substance.Use appropriate work practices when handling to avoid skin or eye contact,ingestion or inhalation.
GC/MSANALYTICAL DATA
Part number(batch number): 30907YB/20245-016-1
Name: Aromatics-Mix2
Column: DB-5ms30mx0.32mmx0.25um
Injection Temperature: 260℃
Inj-vol: 1.00uL
Flow rate: 1.88mL/min
Carrier gas: He
Oven Temperature 60℃(1min),25C/min→210℃(2min),8℃/min→314℃(3min)

注:以上信息仅供参考,以产品附带证书为准。